Polyethylene in knee arthroplasty: A review.
Chakrabarty G, Vashishtha M, Leeder D. Polyethylene in knee arthroplasty: A review. J Clin Orthop Trauma.2015 Jun;6(2):108-12.
Abstract
Background
Polyethylene (PE) has been used extensively in knee arthroplasty since the mid 20th century. Progress in material manufacturing and processing has led to newer polyethylenes over last few decades with different material properties. It has been established that PE wear in knee arthroplasty causes particle induced osteolysis which is the main reason for late failure and requires revision surgery. Although there are various causes of wear, the properties of PE have long been a matter of investigation as a contributory factor. The advent of newer highly cross linked PE has been shown to improve wear rates in hip arthroplasty but the benefits have not been shown to be of the same degree in knee arthroplasty. The laboratory and clinical studies so far are limited and slightly conflicting in their conclusions. The risks of using highly cross linked PE in knee arthroplasty include tibial post fracture, disruption of locking mechanism, liner fracture which can lead to increased wear and osteolysis. The current evidence suggests that highly cross linked polyethylenes should be used with caution and only considered in younger active patients. The results of a recently completed randomized trial to compare the conventional with high molecular weight PE in knee arthroplasty are awaited.
KEYWORDS
Highly cross linked; Polyethylene; Total knee replacement.
Download Full Text Here
For more such Information and Updates, Join the most unique platform in Orthopaedics
and Receive free monthly Newsletter 'Ortho Mirror'
JOIN IORG - CLICK HERE